International Journal of Industrial Biotechnology and Biomaterials

Calibration of Temperature Sensor -Thermocouple and Thermistor in Food Processing Retorts

  1. Kirubanandan Shanmugam ✉

Abstract

The given thermocouple and thermistor are calibrated using stirred liquid bath method. The plot of voltage vs. temperature is drawn for the thermocouple and it is almost nonlinear and consistent with theoretical polynomial equation. Further, it is observed that there is no significant hysteresis of the temperature range of 0°C and 100°C. The nonlinear curve for the thermistor is obtained when the resistance against temperature is plotted. The constant β in the thermistor equation is found to be 3761K. Further, the sensitivity of the given thermistor is 1.00464 which is almost equal to the value of Load resistor’s resistance.

Keywords

References (8)

  1. Bently, E.Robin. (1998) Theory and practice of thermoelectric thermometry: Handbook of Temperature measurement. Springer-Verlag Singapore Pvt Ltd.
  2. T.Gill. (2013) Calibration of Temperature sensors – Thermocouples and thermistors, Food Processing (FOSC 6325) Laboratory handout.
  3. Kevin, B. (2008) Themistors and Thermocouples: Matching the tool to the task in thermal validation, Journal of validation Technology.
  4. Practical Temperature measurements, (2000) Agilent Technologies.
  5. Manual on the use of thermocouples in temperature measurement, (1993) ASTM international.
  6. Web sources - http://www.cantherm.
  7. com/products/thermistors/choosing_ntc.html#app3
  8. Cantherm (2012) How to choose and work with NTC thermistors. http://www.cantherm.com/products/thermistors/choosing_ntc.html#app3